Key Features
- power in AC : 12 kV AC, 40 mA
- power in DC : 12 kV DC, 10 mA
- RS232, IEEE488-2, PLC or 0-10V interfaces
- multi-lingual software
- easy to use and user friendly interface
- dedicated for laboratory and production lines
Downloads
Please feel free to download any of the documents associated with this product.
Technical Specification
DIELECTIC STRENGTH TEST FUNCTION
Output Voltage
- 0 to 12 kV AC
- Accuracy in AC: ± (3% + 50V) of the preset value between 100 to 12000V and for a current <1mA with the detection modes: ΔI, IMAX or ΔI+IMAX
- 0 to 12kV DC
- Accuracy in DC: ± (3% + 100V) of the preset value between 100 to 12000V and for a current <0.5mA with the detection modes: ΔI, IMAX or ΔI+IMAX
Voltage Reading
- On a digital kilovoltmeter connected on the output terminals
- Accuracy ± (1.5% + 20V)
- Display: 1200 digits
Stability
- Less than 1% for a mains variation of ± 10%
Current
- Short-circuit current: 50mA AC
- Short-circuit current: 10mA DC
Current Reading
- On a shunt resistor inserted in the test circuit
- Accuracy: ± (2.5% + 20U) with 1U = 0.01mA
- Display 1000 digits
Breakdown Detection in AC
- "DELTATEST" detector adjusted for ΔI = 1mA ±10% with 10μsec. ±20%
- Total insensitivity to current due to the resistance and the capacitance of the device under test
- "IMAX" detection by maximum current adjustable from 0.01mA steps
- DELTATEST and IMAX mode combination
Breakdown Detection in DC
- "DELTATEST detector adjusted for ΔI = 1mA ±10% with 10μsec. ±20%
- Total insensitivity to current due to the resistance and the capacitance of the device under test
- "IMAX" detection by maximum current adjustable from 0.1mA steps
- DELTATEST and IMAX mode combination
IMIN Threshold
- Adjustable from 00.0mA to 10.0mA in DC
- Adjustable from 00.0mA to 40mA in AC
Breakdown Indication
- By visual (LCD screen and LED) and sound signal
- Breakdown voltage and current are stored on the LCD display
- HV primary transformer shorted when the output voltage is switched off
Timer
- Rise, hold and fall time adjustment between 0 and 999sec.
Storage
- 10 test parameter (voltage, threshold, time...) sets can be stored
View full technical specification for the RMG12AC-DC hipot tester
Product Matrix
The Product Matrix allows you to compare the features of up to 5 products. Simply check the box next to the products you wish to compare then select the 'Compare Products' button.