RXS50/56 Hipot Tester
The RXS50 is a hipot tester designed for use on production lines or in laboratories
▪ AC/DC dielectric strength test (DC in option)
▪ RS232 or IEEE488-2 multi-lingual software
▪ easy to use and user friendly interface
▪ dedicated for laboratory and production lines
▪ complies with the following safety testing standards : EN61010-1, EN60065, EN60335-1, EN60598-1, EN60601-1, EN60950, CEI990, and many others
View RXS50 hipot tester on the manufacturer's website
Key Features
- Suitable for the Aerospace Industry.
- Dielectric Strength - 0 to 5 kV ac - 0 to 6 kV dc 10mA short circuit current.
- Short circuit detection modes: I max, Delta I, BURNING.
- Memorization of 20 parameter settings.
- RS232, IEEE-488, PLC interface
- Burning Mode.
- Voltage ramp facility with timer - from 0 to 999 sec.
Downloads
Please feel free to download any of the documents associated with this product.
Technical Specification
DIELECTRIC STRENGTH TEST FUNCTION
Output voltage
- 0 to 5 kV AC (50 or 60 Hz).
Limited to 4.2 kV AC with FMG501
- 0 to 6 kV DC (RXS56, RXS506)
- Accuracy: ± (2%+50 V) (RXS50) and ± (3%+50 V) (RXS500) of the preset value between 100 to 5000 V and for a current < 100 μA (RXS50) and < 1 mA (RXS500) with the detection modes: DI, IMAX or DI+IMAX
Voltage reading
- On a digital kilovoltmeter connected on the output terminals
- Accuracy: ± (1.5% + 20 V)
- Display: 600 digits
Stability
- Less than 1% for a mains variation of ± 10% (RXS50)
- Less than 3% for a mains variation of ± 10% (RXS500)
Current
- Short circuit < 13 mA AC (RXS50/56) and < 9 mA DC (RXS56) for the max. voltage adjustment
- Short circuit > 200 mA AC (RXS500/506) and >20 mA DC (RXS506) for the max voltage adjustment
- Short circuit duration limited to 5 seconds
- Nominal current : 10 mA AC (RXS50/56) and 4 mA DC (RXS56)
- Nominal current : 10 mA DC (RXS506) and 110 mA AC (RXS500/506)
Current reading
- On a shunt resistor inserted in the test circuit
- Accuracy: ± (2.5% + 2U)
1U = 0.01 mA (RXS50)
1U = 0.1 mA (RXS500)
- Display: 1000 digits
Breakdown detection
- "DELTATEST" detector adjusted for DI = 1 mA ± 10% (RXS50) and DI = 10 mA
± 10% (RXS500) with 10 μsec. ± 20%.
Total insensitivity to current due to the resistance and the capacitance of the device under test
- "IMAX" detection by maximum current adjustable from 0.01 to 10 mA, by 0.01 mA steps (RXS50) and from 0.1 to 110 mA by 0.1 mA steps (RXS500)
- DELTATEST and IMAX mode combination
IMIN threshold function
- Detects whether the probe is properly
connected to the specimen under test
- Adjustable from 0.01 to 10 mA (RXS50) and from 0.1 to 110 mA (RXS500)
DC voltage (RXS56, RXS506 models)
- Positive pole grounded
- Ripple < 1% for I < 100 μA (RXS56)
Breakdown indication
- By visual (LCD screen and LED) and sound signal
- Breakdown voltage and current are stored on the LCD display
- HV primary transformer shorted when the output voltage is switched off
Timer
- Rise, hold and fall time adjustment between 0 and 999 sec.
- Fast mode : (rise + hold) < 900msec.
Storage
- 10 test parameter (voltage, threshold, time,...) sets can be stored
GENERAL CHARACTERISTICS
Presentation
- Table top unit
- Metal case
Dimensions
- Height : 131 mm
- Width : 440 mm
- Depth : 450 mm
Weight
Power
- 230 V or 115 V ± 15% single phase, 4763 Hz
- Consumption: 70 to 600 VA depending on test
Operating temperature
Storage temperature
Overvoltage category
Pollution degree
Safety class
- Class I (earth connection)
View full technical specification
Product Matrix
The Product Matrix allows you to compare the features of up to 5 products. Simply check the box next to the products you wish to compare then select the 'Compare Products' button.