.jpg)
The calibration Standard D05000-CS is designed to enable the full calibration of the DO5000 digital microhmmeter, The resistance ranges are adjusted to a nominal accuracy and the actual value is measured and recorded on the calibration certificate. The value on the calibration certificate is the value that should be used when calibrating the D05000.
Suitable for the Aerospace Industry.
|
CURRENT RANGES |
|
|
Current |
Nominal Accuracy |
|
100mA |
±0.05% |
|
1mA |
±0.05% |
|
10mA |
±0.05% |
|
1mA |
±0.05% |
|
10mA |
±0.05% |
|
RESISTANCE RANGES |
||
|
Resistance Range |
Nominal Accuracy |
Certified Accuracy |
|
30 kΩ |
± 0.01% |
±0.01% |
|
3 kΩ |
± 0.01% |
±0.01% |
|
300 Ω |
± 0.01% |
±0.01% |
|
30 Ω |
± 0.01% |
±0.01% |
|
3 Ω |
± 0.05% |
±0.01% |
|
300 mΩ (200mΩ) |
± 0.05% |
±0.01% |
|
30 mΩ |
± 0.05% |
±0.01% |
|
3 mΩ |
± 1% |
±0.02% |
|
TEMPERATURE RANGES |
||
|
Temperature |
Nominal Accuracy |
Certified Accuracy |
|
0°C |
± 1% |
±0.05% |
|
100°C |
± 1% |
±0.05% |
Working Temperature
0 to +40°C
Storage Temperature
0 to +40°C
Dimensions
460mm x 420mm x 180mm
Weight
5Kg
Terminals
4mm safety sockets. Low thermal EMF types.
Switches
Combination switch with low thermal contacts for the potential selection and low resistance contacts for the current and range selections